B Braun Vista Basic Large Volume Infusion Pump

$1,422.00

In stock

Compare
SKU: 637-202 Category:

Note: This item can only be shipped inside the US and to a medical or educational facility.



The Vista basic was designed to ensure simplicity in all aspects of its use: from set installation, to programming, to special mode use and alarm correction. Based on a design that has sold over 100,000 units worldwide, the Vista basic provides high reliability packaged in one of the easiest pumps to use.



Features and Benefits

  • Robust design of mechanical components ensures maximum durability
  • Two-year preventive maintenance schedule helps save and money from typical annual inspection requirements
  • Battery life of approximately 3 years before replacement is required
  • DEHP-free sets protect patients. Needle-free sets protect both the patient and the clinician
  • Helpful prompts walk the user through the programming process
  • Easy to use, teach and learn
  • Perfect for home care settings








Materials Included

  • Vista Basic Large Volume Infusion Pump
  • (1) Power Cord
  • (1) Mounting Clamp for IV Pole
  • (1) Rechargeable Battery
  • User Manual

Technical Specifications

  • Dimensions
    • Width: 5.5′
    • Height: 9.4′
    • Depth: 7.8′
  • Mechanism: Linear peristaltic volumetric infusion pump
  • Weight: 6.8 lbs. (3.1 kg)
  • Power Requirements: 100-120 VAC, 50-60 Hz, 13 Watts
  • Leakage Current: Max 30A
  • Battery: 7.2V, 1.2 Ah NiCd (rechargeable)
  • Battery Operating Time: 3.5 hours at 125 mL/hr
  • Rate Range: 0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™ tubing
  • Volume to be Infused Range: 0.1 to 9999.9 mL
  • Occlusion Pressure
    • Variable Settings
      • Low: 8 PSI
      • High: 17 PSI
  • KVO Rate: 1 mL/hr or selected rate, whichever is lower
  • Flow Rate Accuracy: ± 5% measurement time 8hr, temp. 71.6° F
  • Alarms
    • Air-in-Line
    • Low Battery
    • Plug In Now
    • Bag Empty
    • Downstream Occlusion
    • Door Open
    • Pump Stopped
    • KVO
    • Unit Malfunction
  • Air-in-Line Alarm
    • Air bubble size 0.3mL* or 1.5mL/hr** (*Can be set from 0.01 to 0.3 mL in service program, **Can be set from 0.5 to 3.5 mL/hr in service program)
  • Memory
    • User option to retain last Continuous and Piggyback Mode data upon power up. Permanent data retention in Program and Ramp/Taper Modes after pump is turned off
  • Tubing: B. Braun Vista™ administration sets
  • Modes/Features: Continuous and Piggyback Modes, Program Mode for variable infusions, Ramp/Taper Mode for TPN infusions and Delay A Mode for postponing infusionstart and finish
  • Operating Temperatures
    • Operating: 64.4° F to 95° F, 10.15 PSI to 15.37 PSI
    • Non-Operating: -13° F to 131° F, 10.15 PSI to 15.37 PSI

Warranty Information

  • 1-Year Warranty

Documentation

User Manual

Quick Reference Guide

Main Menu